Set-up

The unit can be configured to suit most applications. Additional hoses, fittings, and adapters
may be required for particular uses.
1. For use with external oxygen: connect a green (white for some installations) high-pressure
oxygen hose to the OXYGEN IN fitting and a 55 psig external source.
Note:
Connect the disposable ventilator circuit to the GAS OUTPUT, TRANSDUCER, and
2.
EXHALATION VALVE connectors on the unit's Connector Panel. Follow the directions
included with the disposable ventilator circuit.
In a high-dust or biological environment, a bacterial/viral filter should be attached to the
3.
Fresh Gas/Emergency Air Intake to prevent entrainment of particulate or biological matter.
(See "Hazard Environment Filters" section.)
4.
In toxic biological or chemical environments, the operator can attach a chemical/biological
filter to the Fresh Gas/Emergency Air Intake. The threaded interface accommodates
chemical/biological filters with an Rd 40 x 1/7 interface (see BS EN 148-1 1999 Respiratory
protective devices-threads for face pieces). (For more information, see the "Hazard
Environment Filters" section of this manual.)
For use with AC power: connect the AC/DC Power Supply (supplied) to the External Power
5.
Input.
Note:
Warning!
Always assure that there is an alternate means of providing ventilation. A bag-valve
resuscitator and an appropriate mask for the patient being ventilated should be
immediately available.
Warning!
This unit's USB connection does not provide any signal output or input to the operator.
It is for the use of ZOLL-trained personnel only. The operator should not connect
anything to the USB connection.
906-0731-01 Rev. J
Use only with medical-grade (USP) oxygen. When using with an oxygen cylinder,
the cylinder must be secured.
The unit can operate from internal batteries or from external AC or DC power
sources.
ZOLL Portable Critical Care Ventilator