ZOLL 731 Series Operator's Manual page 126

Portable critical care ventilator
Hide thumbs Also See for 731 Series:
Table of Contents

Advertisement

S
PECIFICATIONS
The unit was tested for the effect of pressure at the patient connection port (e.g airway pressure)
on the V
The tests consisted of a core matrix of volume or pressure targets for a set of resistances and
compliances, based on the ASTM F1100 guidance. These test cases were then performed at
different PEEP targets, FIO
Gas Intake accessories. The possible settings and accessories are listed below:
Setting/Accessories
PEEP Settings
FIO
Settings
2
Patient Circuit Accessories
Intake Filter Accessories
The following accuracies/precision were observed for adult settings with non-MRI breathing
circuit:
Parameter
V
, Tidal Volume
T
Pressure Target Accuracy
PEEP Accuracy
FIO
Accuracy
2
A-8
and Oxygen concentration.
T
settings, and with different patient circuit accessories and Fresh
2
www.zoll.com
Value
0, 5, 25
21, 30, 50, 90, 100
No Accessory
HME
HME + BV
HMEBV
No Accessory
C2A1
BV Filter
O2 Reservoir
Accuracy/Precision
±
10% of setting
±
5 cm H
O for setting < 30, else
2
±
±
1.5 cm H
O
10 of setting
2
±
±
3% of full scale
10 of setting
±
10 cm H
O
2
906-0731-01 Rev. J

Advertisement

Table of Contents
loading

Related Products for ZOLL 731 Series

This manual is also suitable for:

Eagle iiEmv+Aev

Table of Contents