LARMS
Service Code
Alarm Name/Mitigation/Resolution
1041
O2 Supply Pressure High
The oxygen supply is >80 psig. Pressures above 80 psig could result in a catastrophic failure, harm
to the patient and/or damage to the unit.
Mitigation/Info: Manually Ventilate Patient, Decrease O2 Supply to 55 psig, Replace Regulator,
Connect Low Flow O2
**Restart Ventilator without O2 Supply**
1051
Run-Time Calibration Failure
There is a failure of the calibration system.
Mitigation/Info: Manually Ventilate Patient
**Replace/Service Ventilator**
1052
Airway Pressure Sensing Failure
Communication between the airway pressure sensor and SPM is lost.
Mitigation/Info: Manually Ventilate Patient
**Replace/Service Ventilator**
1061
Exhalation System Failure (Excessive Airway Pressure)
The airway pressure (P
H
O) for >5 seconds, or when the P
2
the unit stops ventilating and attempts to discharge the pressure in the breathing circuit to
atmosphere.
Mitigation/Info: Manually Ventilate Patient, Replace Circuit and Restart
**Replace/Service Ventilator**
1061
Exhalation System Failure (Excessive Airway Pressure)
The airway pressure (P
H
O) for >5 seconds, or when the P
2
the unit stops ventilating and attempts to discharge the pressure in the breathing circuit to
atmosphere.
Mitigation/Info: Manually Ventilate Patient, Replace Circuit and Restart
**Replace/Service Ventilator**
1172
5 Volt Self Check Failure
The 5 volt power bus fails to provide the required voltage.
Mitigation/Info: Manually Ventilate Patient, Ventilator Not Functioning
**Replace/Service Ventilator**
1173
Internal Communication Failure (Host Device Communication Failure)
Communication fails between one of the subcomponents and the hose processor.
Mitigation/Info: Manually Ventilate Patient, Backup Ventilator Enabled
**Replace/Service Ventilator**
1174
Off Set Self Check Failure
The device is not able to calibrate the one or more transducers and is no longer able to operate
safely.
Mitigation/Info: Manually Ventilate Patient, Ventilator Not Functioning
**Replace/Service Ventilator**
4-12
) is above 40 cm H
O, the PIP High Limit (when PIP High Limit is <35 cm
2
aw
is above 75 cm H
aw
) is above 40 cm H
O, the PIP High Limit (when PIP High Limit is <35 cm
2
aw
is above 75 cm H
aw
www.zoll.com
O for >1.5 seconds. When this happens,
2
O for >1.5 seconds. When this happens,
2