ZOLL 731 Series Operator's Manual page 102

Portable critical care ventilator
Hide thumbs Also See for 731 Series:
Table of Contents

Advertisement

A
LARMS
Service Code
Alarm Name/Mitigation/Resolution
1060
Exhalation System Failure (Exhalation Valve Failure)
The exhalation control valve fails to operate. When this happens, the unit stops ventilating and
attempts to discharge the pressure in the breathing circuit to atmosphere. This failure may be
caused by a significant blockage of the exhalation valve or an occlusion/kink in the exhalation valve
tube.
Mitigation/Info: Manually Ventilate Patient, Replace Circuit and Restart
**Replace/Service Ventilator**
4-24
www.zoll.com
906-0731-01 Rev. J