ZOLL 731 Series Operator's Manual page 61

Portable critical care ventilator
Hide thumbs Also See for 731 Series:
Table of Contents

Advertisement

Pop Up message will appear and ask for confirmation.
5.
Press Accept.
6.
Once the patient has reached steady state, the alarms should be configured to closely
7.
monitor the patient.
To start NPPV in AEV
1. Press MODE button to highlight mode.
2.
Adjust to CPAP.
Pop Up message will appear and ask for confirmation.
3.
Press Accept.
4.
5.
Once the patient has reached steady state, the alarms should be configured to closely
monitor the patient.
When the NPPV adjunct is in operation, the head with mask icon will appear in the location
used by the speaker/mute icons. Low and Medium priority alarms will cause this head with
mask icon to disappear. It will reappear when Low priority alarms are muted. When Medium
priority alarms are muted, the muted speaker icon will appear.
When transitioning to NPPV mode, the following parameters/alarm limits will automatically be
adjusted to the following default values:
Alarm/Parameter
High BPM Alarm
Low BPM Alarm
Low Airway Pressure Alarm
PEEP
V
High Limit
T
V
Low Limit
T
Rise Time
Pressure Support
When transitioning from the NPPV mode to another mode, the following parameters/alarm
limits will reset to the default values. This is done to assure appropriate monitoring in an active
ventilating mode:
Alarm/Parameter
High BPM Alarm
Low BPM Alarm
Low Airway Pressure Alarm
PEEP
V
High Limit
T
906-0731-01 Rev. J
®
:
CPAP into NPPV Default Value
40
4
Off
5 cm H
O
2
Off
Off
3
2 cm H
O
2
NPPV into CPAP Default Value
25
8
(PIP Target = -3)
No change
750
ZOLL Portable Critical Care Ventilator
NPPV to P Default Value
25
8
(PIP Target = -3)
No change
750
3-5