Device Alarms

Device alarms are triggered by a malfunction or internal fault with the ventilator device.
Service Code
Alarm Name/Mitigation/Resolution
1001
Compressor Failure (Compressor Control Fault - No Backup)
The compressor fails to operate or fails to provide the flow required to deliver a breath and
high-pressure oxygen (HP O
Mitigation/Info: Manually Ventilate Patient, Connect HP O
**Replace/Service Ventilator**
1002
Compressor Failure (Compressor Signal Chain Fault - No Backup)
Communication between the compressor controller and Smart Pneumatic Module (SPM) is lost and
high-pressure oxygen (HP O
Mitigation/Info: Manually Ventilate Patient, Connect HP O
**Replace/Service Ventilator**
1003
Self Check Failure
The flow from the first breath is ±20% of the expected flow for the tidal volume at start up. This
unusually low RPM is a symptom of a dirty flow screen which cannot be serviced by the operator.
Mitigation/Info: Manually Ventilate Patient
**Replace/Service Ventilator**
1010
O2 Valve Failure (O2 Valve Failed Open)
The O2 Valve fails in the open position which results in continuous inspiratory flow. When this
occurs, the unit automatically opens the exhalation valve to prevent pressure from accumulating in
the circuit and ventilation stops.
Mitigation/Info: Manually Ventilate Patient
**Replace/Service Ventilator**
1011
O2 Valve Failure (O2 Valve Control Fault - No Back Up)
The signal to the O
available to provide ventilation.
Mitigation/Info: Manually Ventilate Patient
**Replace/Service Ventilator**
1012
O2 Valve Failure (O2 Valve Signal Chain Fault - No Back Up)
The communication between the O
provide ventilation.
Mitigation/Info: Manually Ventilate Patient
**Replace/Service Ventilator**
1020
O2 Supply Pressure Low
The oxygen supply pressure is
If the oxygen source can be restored, the unit should be cycled off then on to reset. By design, the
unit will not reestablish oxygen operation unless the supply pressure is
Mitigation/Info: Manually Ventilate Patient, Connect 55 psig O2 then Restart, Check O2 Supply for
Leaks,Replace Regulator
**Replace/Service Ventilator**
906-0731-01 Rev. J
) is not available to provide ventilation.
2
) is not available to provide ventilation.
2
Valve is not delivering the required flow rate and the compressor is not
2
Valve and the SPM fails and the compressor is not available to
2
35 psig and the compressor is not available to support ventilation.
ZOLL Portable Critical Care Ventilator
, Restart Ventilator With HP O
2
, Restart Ventilator With HP O
2
40 psig.
2
2
4-11