description of how to enter the NPPV (Noninvasive Positive Pressure) adjunct in the Mode
parameter window. To access the Mode Context menu, press and hold the MODE parameter
button. This will permit the Apnea Back Up settings to be set.
Menu Item
BPM
PIP Target
I:E, TI
When going into Apnea Back Up from CPAP in the EMV+
PPV), and in the AEV
following back up settings:
PIP Target
20
All other parameters will remain as they were previously set. When leaving CPAP, the new
operating mode will be Pressure Targeted and will use Apnea Back Up settings as the new
settings.
Caution
These Apnea Back Up settings are appropriate for an adult. For pediatric and infant patients, the
operator should adjust the Apnea Back Up settings so they are appropriate for the patient.
Confirm/Select -- Press the CONFIRM/SELECT button to confirm a new control setting
8.
or to select from a menu or setting option. The CONFIRM/SELECT button switch is
labeled with a green check mark
9.
Power Off/On -- Turn the POWER OFF/ON switch to apply or remove operating power to
the unit.
Manual Breath -- Pressing the MANUAL BREATH button delivers one control breath
10.
based on the V
inspiration or before baseline is reached in the expiratory period, a manual breath will not be
triggered. If there is incomplete exhalation, it will not be possible to trigger a manual breath.
This is to prevent accidental breath stacking. When operating in CPAP, a breath is delivered
using the PIP value from the Apnea Back Up settings.
Rotary Encoder -- Turn the ROTARY ENCODER to change a value or highlight a
11.
particular menu option.
Mute/Cancel -- Press the MUTE/CANCEL button to mute most Medium Priority alarms, to
12.
cancel/acknowledge Low Priority alarms, or to cancel an action that is no longer desired (for
example: a parameter value change). Pressing the MUTE/CANCEL button can also be used
to cancel any current operation and return to the normal operating screen.
The MUTE/CANCEL button switch is labeled with a red "
Menu -- Pressing the MENU button provides access to user menus and special functions.
13.
Use the ROTARY ENCODER to scroll to the desired menu option and press the
CONFIRM/SELECT button to access the menu control. Pressing CONFIRM/SELECT
then accesses the parameter variable which is changed by turning the ROTARY ENCODER
to the desired value. To accept the new parameter value, press the CONFIRM/SELECT
906-0731-01 Rev. J
Mode Context Menu
Range
1 to 60
10 to 80
1:1 to 1:99, 0.3 to 3
®
(NPPV only), the unit will automatically return to Pressure with the
BPM
12
or PIP settings. If the MANUAL BREATH button is pressed during
T
ZOLL Portable Critical Care Ventilator
Description
Allows operator to select BPM for Apnea Back
Up
Allows operator to select the PIP Target for
Apnea Back Up
Allows operator to select either I:E or TI for
Apnea Back Up
1:2.5 OR 1.42
.
"
"
®
and Eagle II
(NPPV and
I:E OR T
I
"
X
2-13