ZOLL 731 Series Operator's Manual page 36

Portable critical care ventilator
Hide thumbs Also See for 731 Series:
Table of Contents

Advertisement

P
O
RODUCT
VERVIEW
(the highlight moves from the parameter variable back to the parameter label). For
parameters with multiple options, pressing CONFIRM/SELECT opens a submenu where
the various parameters are selected using the ROTARY ENCODER and changed using the
process described above. At any point, the operator can cancel an operation, return to the
previous menu level, or exit the menu control by pressing the MUTE/CANCEL button.
1. Alarm Configuration -- allows the operator to disable the audible and visible alarm
alerts associated with specific alarm in order to prevent nuisance alarms. While the
menu shows all of the alarms that can be disabled, the system only allows you to disable
active alarms. This approach assures that the operator has been advised of the alarm and
has decided that it does not affect safe and effective management of the patient. It is
important to recognize that Low Priority alarms are Advisory and the patient and device
are performing as configured, but there is information that could affect management of
the patient.
Power Up Settings -- allows the operator to select startup settings different from the
2.
factory defaults. The menu allows the operator to select the current ventilator settings for
use at startup. There are 3 options: default, user 1, and user 2. User 1 and user 2 allow the
operator to establish different startup settings based on the intended patient population.
ZOLL is constantly focused on providing clinically safe, effective and timely support to
patients requiring ventilator support as well as ease of use to their care providers. The
initial focus and area for application of the unit platform is Acute Patient Care in many
varied and austere environments. This patient area of need has historically been the
Adult range of patients. ZOLL has used this historical data and experience as
its rationale for the unit to start-up in the Adult patient Assist-Control Volume mode and
associated parameters which have been deemed as clinically safe and acceptable starting
points and limits by international experts and standards committees.
ZOLL also strives to provide user-friendly options to our customers and allow them to
set up certain USER Start-up selected modes after their acquisition of the equipment as
they know best what their patient base and needs for mechanical ventilatory support are
within their respective region or facilities. Realizing that varied levels of need and
experience exist, as well as knowing that each patient is different and clinician
assessment is required prior to application to the patient, certain mode and parameter
configurations are not available as User setup options in an effort to maximize patient
safety.
To configure startup settings other than the default, press the MENU button, select
POWER UP SETTINGS; select SAVE SETTINGS and select either user 1 or user 2
and press CONFIRM/SELECT. To start with the new setting, select POWER UP
WITH and select which user setting you would like to use at startup and press
CONFIRM/SELECT. Assure the proper startup setting by turning the ventilator off
then on again. The unit will being operation with the new settings. Limits for stored
parameters are shown on the table below. Default startup modes are limited to AC and
SIMV (SIMV available for EMV+
stored default.
Parameter
V
V
V
Constant I time vs. constant I:E
2-14
Setting
T
High Limit
T
Low Limit
T
www.zoll.com
®
and Eagle II
Limitations
≤1000 ml
10ml resolution, only recalled if the target is pressure or in SIMV(V)
10ml resolution, only recalled if the target is pressure or in SIMV(V)
No limit
only); CPAP cannot be stored as a
906-0731-01 Rev. J