Alarm Turn Off At Extreme Range Limits - ZOLL 731 Series Operator's Manual

Portable critical care ventilator
Hide thumbs Also See for 731 Series:
Table of Contents

Advertisement

A
LARMS
Alarm
Number
2411
3300
3301
3310
3311
3312
3313
3315
3316
3317
3318

Alarm Turn Off at Extreme Range Limits

If the operator sets the following alarm limits to their extreme range, the indicated alarms will
be turned off after the Pop Up message confirmation:
1.
2.
a.
3.
a.
4.
a.
5.
6.
7.
If an alarm has been turned off and is then modified, but is not accepted, then the alarm
parameter will be set to the values indicated in the following table. This is done to ensure
patient safety in the event of an inadvertent value change. These values can be changed by the
operator following the parameter change procedures described above.
High Breath
PIP Low
Rate
99 BPM
3 cm H
O
2
4-26
Alarm Name
Heart Rate Low (Pulse Rate Low)
SpO2 Shutdown (MS 11 Failure-Monitor Not In Use)
SpO2 Shutdown (Communication Failure EMV-Pulse Ox-Monitor Not In Use)
No SpO2 Sensor Connected (No Sensor Detected)
Defective Sensor
SpO2 Pulse Search
SpO2 Signal Interference
Too Much Ambient Light
Invalid SpO2 Sensor (Unrecognized Sensor)
Low SpO2 Perfusion (Low Perfusion)
Low SpO2 Perfusion (Poor SpO2 Signal
High Breath Rate (Alarm #2074)
PIP Low (Alarms #2071, 2171)
This changes automatically to off in NPPV.
V
High (Alarms #2072, 2172)
T
This changes automatically to off in NPPV.
V
Low (Alarms #2073, 2173)
T
This changes automatically to off in NPPV.
Low SpO2 (Alarm #2410)
High Heart Rate (Alarm #2410)
Low Heart Rate (Alarm #2411)
V
High
T
2000 ml
0 ml
www.zoll.com
V
Low
Low SpO2
T
86%
High Heart
Low Heart
Rate
Rate
240 BPM
30 BPM
906-0731-01 Rev. J