ZOLL 731 Series Operator's Manual page 41

Portable critical care ventilator
Hide thumbs Also See for 731 Series:
Table of Contents

Advertisement

M. Battery Icon/Indicator -- Indicates (1) the presence of a functional battery, (2) when the
battery is charging, and (3) the current battery capacity. The Battery icon appears in outline
form and is filled with horizontal lines indicating its current capacity. When the battery is
charging, these horizontal lines cyclically scroll vertically, one row at a time, from the
bottom to the top. The scroll rate is faster when the battery is charging faster and it slows
down when slower trickle charging. The charge level is indicated numerically under the
battery icon. When the battery is fully charged, the icon is completely filled and scrolling
stops. Each line represents approximately 5% of battery capacity. During internal battery
operations, a horizontal line is removed with each 5% decrement of battery capacity. The
Battery icon will flash off/on when a Battery Power Low alarm occurs. The icon will flash
off/on and present with a diagonal line when no battery is connected.
N. Oxygen Supply Icon/Indicator -- Indicates the presence of external high-pressure oxygen
(55 psig source). The icon only appears when high-pressure oxygen is detected by the
pressure transducer. The icon flashes off/on when the Oxygen Low/Fail alarm occurs.
O. Speaker (Mute or Unmute) Icon/Indicator -- Indicates whether the alarm sound has been
muted or is active. The Muted speaker icon is present during the startup 2 minute mute, with
the remaining muted time indicated under this icon.
O. Head with Mask Icon/Indicator -- Indicates if the NPPV adjunct is in effect for ventilatory
support of spontaneously breathing patients. This icon replaces the Unmuted Speaker and/or
the Muted Speaker icon for Low priority alarms. (This icon does not appear if a Medium
priority alarm has been muted.)
P. Alarm Triangle Icon/Indicator -- Indicates if there is an active alarm. The priority of the
alarm is indicated by the interior of the triangle: empty for Low priority, one exclamation
point for Medium priority, and two exclamation points for High priority.
Q. Airway Pressure/Measured Values -- Provides a continuous display of airway pressure. Its
absolute range is from -10 to 100 cm H
H
O/pixal. The scale below the indicator is graduated in 10 cm H
2
numerical markers appearing at 0, 50, and 100 cm H
values for delivered minute volume, mean airway pressure (MAP), and actual breathing rate
(set plus spontaneous) are displayed.
906-0731-01 Rev. J
O ATPD with a horizontal resolution of 1cm
2
ZOLL Portable Critical Care Ventilator
O increments with
2
O. Above the bar graph, measured
2
2-19

Advertisement

Table of Contents
loading

Related Products for ZOLL 731 Series

This manual is also suitable for:

Eagle iiEmv+Aev

Table of Contents