Service Code
Alarm Name/Mitigation/Resolution
2053
Suspicious Triggers
Airway pressure sensor fails to calibrate the expiratory phase of a breath. When this occurs, the
unit attempts to reestablish a baseline by momentarily setting PEEP to 0 cm H
triggered breaths. This interruption lasts no longer than 2 breath cycles.
Mitigation/Info: Attempted Self-Calibration, Momentarily Disabling Triggers and PEEP, Check
Circuit for Leak/Disconnects, Check Tube Placement/Cuff
**Replace/Service Ventilator**
2062
Exhalation Fault
The airway pressure (P
pressure (PEEP pressure).
Mitigation/Info: Check Circuit for Kinked Hose/Tube, Check for Blocked Exhalation Valve, Replace
Circuit, Replace/Service Ventilator
**Manually Ventilate Patient**
2070
Airway Pressure High
The P
aw
this occurs, flow decelerates to maintain P
the breath (inspiratory time).
Mitigation/Info: Pressure Exceeds Limit Setting, Check Circuit for Kinked Hose/Tube, Check for
Airway Obstructions, Suction Airway if Necessary, PIP High Limit Set Too Low?
**Manually Ventilate Patient**
2071
Low Airway Pressure
The P
aw
Mitigation/Info: Check Patient Connection, Check Circuit for Loose Hose/Tube, Check Exhalation
Valve, Check Tube Placement/Cuff, Is High Limit Set Correctly?
**Manually Ventilate Patient**
2072
High Tidal Volume
Operating with a pressure target and the delivered tidal volume exceeds the operator defined limit
for 2 consecutive breaths.
Mitigation/Info: Check Patient Connection, Check Circuit for Loose Hose/Tube, Check Exhalation
Valve, Check Tube Placement/Cuff, Is High Limit Set Correctly?
**Monitor Patient**
2073
Low Tidal Volume
Operating with a pressure target and the delivered tidal volume does not reach the operator defined
limit for 2 consecutive breaths. When this occurs, flow decelerates to maintain the P
pressure limit for the duration of the breath (inspiratory time). If the PIP setting is set properly, the
breath should be greater than the low limit (provided it is set properly as well).
Mitigation/Info: Check Circuit for Kinked Hose/Tube, Check For Airway Obstruction, Suction Airway
If Necessary, If Low Limit Set Correctly?
**Manually Ventilate Patient**
906-0731-01 Rev. J
) measured at the end of expiration is >5 cm H
aw
meets or is greater than the High Airway Pressure Limit for 2 consecutive breaths. When
is less than the Low Airway Pressure Limit for 2 consecutive breaths..
ZOLL Portable Critical Care Ventilator
at the High Airway Pressure Limit for the duration of
aw
O and suspending
2
O above the baseline
2
at airway
aw
4-15