Operating Power Selection And Stopping - ZOLL 731 Series Operator's Manual

Portable critical care ventilator
Hide thumbs Also See for 731 Series:
Table of Contents

Advertisement

Recommended
Patient
Patient Weight
Adult
40
Pediatric
12-50
Pediatric
12-30
Infant
5-12
Warning!
If circuits with different resistance/compliance are used or additional adjuncts are
placed in line with the circuit, the operator must make sure that the appropriate circuit
resistance/compliance factors for the new circuit are used as well as dead space volume
of the added adjunct(s) is/are factored into the equation for effective tidal volume
delivery to the patient.

Operating Power Selection and Stopping

The POWER switch is the master power switch. Use this switch to indicate or end operation.
The unit is designed to operate using any one of three power sources:
1. Internal 14.8V lithium ion (Li Ion) rechargable battery with 6.6 Ah capacity (fully charged,
the battery provides 10 hours of operation at factory default settings with pulse oximeter
operating at 25C).
External AC/DC Power Supply (110-240 VAC 50/60 and 400 Hz--ZOLL verified testing,
2.
and 110-240 VAC 50/60--SGS listed) with IEC 320 style AC input connector (supplied). The
AC/DC Power Supply provides a DC output of 24V at 4.2A.
External DC power from a standard vehicle DC outlet using the 12 or 28 VDC Power Cables.
3.
The input connector of the EMV+ is designed to accept DC voltages between 12.5 to 28.0
VDC +/- 10% volts, negative ground.
Caution
When using the standard vehicle DC outlet, the vehicle should not be jump started during
operation of the ventilator.
Warning!
Never start the ventilator with the patient connected. Always start the ventilator, select
the patient settings, ensure operation, and then connect the patient. Always manually
ventilate the patient when they are not connected to the ventilator.
The unit is designed to use external power when available rather than its internal battery pack.
When an acceptable external power source is present, the internal battery pack is automatically
charged while the unit operates. When an external power failure occurs, the unit automatically
switches to its internal battery pack for operating power and activates the EXTERNAL POWER
FAILURE alarm; there is no interruption in operation and/or loss of any alarms. When external
power returns, operating power automatically switches from internal power to the external
source. In the event that the device needs to be shutdown, the procedure consists of turning the
POWER switch to the OFF ("O") position. If this fails to work or puts the patient or operator at
possible risk, the device should be disconnected for the mains power.
906-0731-01 Rev. J
MRI Longer Circuit Length
Breathing Circuit
(kg)
Pediatric/Adult
Pediatric/Adult
Infant/Pediatric
Infant/Pediatric
ZOLL Portable Critical Care Ventilator
Inspiratory
Expiratory
Resistance
Resistance
(cm H
O)
(cm H
2
2.33
4.44
0.65
2.63
1.28
1.75
0.39
1.25
Compliance
(ml/cm H
O)
2
O)
2
2.1
2.1
0.75
0.75
2-29

Advertisement

Table of Contents
loading

This manual is also suitable for:

Eagle iiEmv+Aev

Table of Contents