The following is an alphabetical list of all Pop Up messages and their meaning:
Pop Up Message
ALARM DISABLE
BPM LIMIT CONFLICT
BPM LIMIT CONFLICT
BPM SETTING CONFLICT
BPM SETTING CONFLICT
CPAP + NPPV
HEART RATE LIMIT CONFLICT
HEART RATE LIMIT CONFLICT
HIGH PRESSURE TARGETING
SETTING
HIGH V
HIGH V
I TIME RANGE EXCEPTION
I TIME RANGE EXCEPTION
I:E RANGE EXCEPTION
I:E RANGE EXCEPTION
LOW BREATH RATE SETTING
PEEP BACKUP SETTING
CONFLICT
PEEP SETTING CONFLICT
PEEP SETTING CONFLICT
PEEP SETTING CONFLICT
PEEP + PS SETTING CONFLICT
PIP LIMIT BACKUP SETTING
CONFLICT
PIP LIMIT CONFLICT
PIP LIMIT CONFLICT
PIP LIMIT CONFLICT
PIP LIMIT CONFLICT
PIP SETTING CONFLICT
PIP SETTING CONFLICT
906-0731-01 Rev. J
LIMIT SETTING
T
SETTING
T
ZOLL Portable Critical Care Ventilator
Explanation
Confirmation required-press accept key to disable alarm
Cannot adjust high limit lower than low limit
Cannot adjust low limit higher than high limit
I Time cannot exceed 3 seconds
I:E > 1:99 not allowed
Some Alarms Disabled! Configure Alarms for Patient!
Cannot adjust high limit lower than low limit
Cannot adjust low limit higher than high limit
Confirmation required-press accept key to exceed 60 cm H
Confirmation required-press accept key for values above
1500ml
Confirmation required-press accept key to allow V
I Time cannot exceed 3 seconds
I Time must be greater than 0.3 seconds
I:E > 1:99 not allowed
Inverse I:E not allowed
Confirmation required-press accept key for values below 6
BPM
Cannot adjust PEEP target to within 5 of backup PIP target
Cannot adjust PEEP target to within 5 of PIP High Limits
Cannot adjust PEEP target to within 5 of PIP target
PEEP + PS cannot be greater than PIP High Limit
PEEP + PS must be
3 cm H
Patient Safety
Cannot adjust high limit lower than backup PIP target
Cannot adjust high limit lower than low limit
Cannot adjust high limit lower than PIP target
Cannot adjust high limit lower than PS + PEEP
Cannot adjust low limit higher than high limit
Cannot adjust PIP target higher than PIP High Limit
Cannot adjust PIP target to less than 5 more than PEEP
> 1000ml
T
O, Adjust Setting to Assure
2
O
2
2-21