Emerson Rosemount Reference Manual page 35

Smart pressure gauge
Hide thumbs Also See for Rosemount:
Table of Contents

Advertisement

Reference Manual
00809-0100-4145
From the HOME screen, enter the Fast Key sequence
Fast Keys
Procedure
1. Select 2: Configure.
2. Select 2: Manual Setup.
3. Select 1: Measurements.
4. Select 1: Dial/Pressure.
5. Select 1: Verify/Calibrate.
6. Select 1: Verify/Calibrate Dial+Digital Pressure.
7. Adjust dial indicator until it points to lower endpoint.
8. Select 5: Save Dial.
4.3.5
Recall factory trim—sensor trim
The recall factory trim—sensor trim command allows the restoration of the as-shipped
factory settings of the sensor trim. This command can be useful for recovering from an
inadvertent zero trim of an absolute pressure unit or inaccurate pressure source.
From the HOME screen, enter the Fast Key sequence
Fast Keys
Procedure
1. Select 3: Service Tools.
2. Select 3: Maintenance.
3. Select 2: Restore to Default Settings.
4. Follow the screen prompts to recall sensor and dial trim.
Emerson.com/Rosemount
The following adjustments are available and can be used to complete the dial
adjustment.
Fine counterclockwise (0.1 percent of Span)
Fine clockwise (0.1 percent of Span)
Coarse counterclockwise (0.3 percent of Span)
Coarse clockwise (0.3 percent of Span)
2, 2, 1, 1, 1, 1
3, 3, 2
Operation and maintenance
September 2019
35