Replacing Forks

Ensure that there is sufficient room to turn patient lift on its side and that floor area
is clear of debris. Otherwise, injury to personnel or damage to patient lift may
occur.
1. Place the patient lift on its side.
2. Remove the front or rear caster from the lift. Refer to Replacing Casters/Forks on 
page 40.
3. Unscrew the existing fork from the base.
4. Install the new fork onto the base.
5. Install the front or rear caster onto the patient lift. Refer to Replacing Casters/Forks on 
page 40.
6. Stand the patient lift back up.
Part No. 1145810
WARNING