Condensate Collecting Tray - Binder VDL Series Mounting Instructions

Vacuum module for vacuum drying ovens
Hide thumbs Also See for VDL Series:
Table of Contents

Advertisement

Available languages

Available languages

GB
GB
5.

Condensate collecting tray

Safety concept:
If solvents are accidentally spilled when removing the filled condensate collector flask from the pump,
they are collected by the tray. Supplementary safety measures are the extraction from the vacuum
module (chap. 7) as well as the prescribed installation of the module on a conductive surface with a
conductive connection of the module, pump, and collecting tray to this surface (chap. 6)
The condensate collecting tray comes with the BINDER pumps VP4 and VP5.
You can also order the condensate collecting tray separately:
Size of vacuum module
23 / 53 / 115
115
The following figures show the condensate collecting tray in blue color.
Vacuum module size 23 / 53 / 115 with vacuum
pump VP4 (MZ 2C EX) and mounted conden-
sate collecting tray
Mounting instructions for vacuum module10/2016
Pump
VP4 (MZ 2C EX)
VP5 (MD 4C EX)
Vacuum module size 115 with vacuum pump VP5
(MD 4C EX) and mounted condensate collecting tray
Collecting tray, Art. no.
4022-0365
4022-0366
Page 18/32

Advertisement

Table of Contents
loading

This manual is also suitable for:

Vd seriesVd 53Vd 115Vdl 23Vdl 53Vdl 115 ... Show all