INTRODUCTION
1.5

Sampling System

The Liquidew I.S. requires a clean sample of the process liquid that meets the
temperature, pressure and fl ow requirements of the transmitter. The design of the
sampling system will depend on the specifi c application.
The requirements for the sample liquid going into the sensor block are as follows:
Temperature: 0 to +40°C (+32 to +104°F) (maximum +60°C (+140°F))
NOTE: Sample temperature must be above the moisture saturation
temperature of the sample fl uid
Maximum pressure: 45 MPa (450 barg / 6526 psig)
Flow rate: 0.1 to 0.3 l/min (0.026 to 0.079 GPM)
NOTE: Contact Michell Instruments if you wish to order a specifi c sampling
system.
Please refer to the Liquidew I.S. Premium Sampling System Instructions if a Michell
sampling system has been ordered with the Liquidew I.S.
8
Liquidew I.S. User's Manual
97092 Issue 5.1, October 2018