B Autofilling accessories

Setting up a Weigh Station

Setting up a Weigh Station
About this topic
Calibrating a Weigh Station
182
This topic provides calibration instructions for the Weigh Station.
For a description of the Weigh Station and installation instructions, see the
Module User
Guide.
Crosshairs for setting
teachpoint
00268
Bravo
weigh pad
Before you begin
Make sure the Pump Module is set up and connected to the Weigh Station. See the
Pump Module User
Guide.
Verify that the Autofill Station accessory is configured in Bravo Diagnostics. See
"Configuring an autofilling station for the Bravo Platform" on page
Procedure
Before the Weigh Station is ready for use in a protocol, you must calibrate the empty
and full settings. When you calibrate the Weigh Station, you must use the reservoir type
that will be weighed.
To calibrate a Weight Station:
1
In Diagnostics, click the Profiles tab and make sure the correct profile is initialized.
2
Click the Configuration tab.
3
In the Accessory Configuration area, highlight Autofill Station, and then click
Diagnose accessory.
Note: The Autofill Station is represented by a Pump Module graphic in the
software.
Pump
177.
Bravo Platform User Guide