IMPORTANT
Figure Vacuum Filtration Station connections for a pinch-valve module
1
Bravo Platform User Guide
5
Optional. Disconnect the manual on/off toggle switch (5) from the end of the
tubing. This toggle switch is not required for the setup.
Connecting the pinch valve tubing
The house vacuum must be regulated to a maximum of 286.49 mbar (8.46 inHg),
because the pinch-valve tubing will collapse with higher vacuum pressure.
Note: The pinch-valve tubing can collapse with regular use over time, resulting in
inadequate vacuum flow for the Vacuum Filtration Station. Replace the tubing
periodically to keep the Vacuum Filtration Station functioning as intended.
The following figure shows the connections from the Vacuum Filtration Station to a
vacuum source via a pinch-valve module.
Item
Name
1
Vacuum control
valve-and-gauge
assembly
2
Pinch-valve module assembly
3-way
connector
Installing and setting up a Vacuum Filtration Station
2
6 7
Description
Helps regulate the vacuum pressure to prevent the
pinch-valve module tubing from collapsing.
Branches the tubing from the Vacuum Filtration Station
base to the vent and valve ports on the pinch-valve
module.