Removing a Pump Module
Related information

Installing a Weigh Station

About this topic
About the Weigh Station
Pump Module User Guide
To remove the Pump Module:
1
Using the automation control software, run the Pump Module and drain
the reservoir until empty.
2
Turn off the Pump Module, and then turn off the liquid- handling device.
3
Disconnect the Pump Module power and communication cable. Unplug the
Weigh Station cable from the Pump Module.
4
Disconnect the tubing from the reservoir.
5
Remove the reservoir.
6
Remove the tubing from the Pump Module and from the source and waste
bottles.
For information about...
Autofilling reservoirs
Configure the Pump Module,
reservoir, and associated resources
Routine maintenance
This topic describes the Weigh Station and how to install and remove it from
the physical workspace.
For details on connecting the cable, see
communication" on page
A Weigh Station measures the weight of a Agilent autofilling reservoir that is
placed on it. When appropriately configured in the software, the Weigh Station
works with the Pump Module to ensure that the reservoir is filled to a
constant liquid level during the pump reagent task in a protocol. By
monitoring the weight of the reservoir that sits on it, the Weigh Station
controls when the Pump Module is activated.
Use one Pump Module for each pairing of reservoir and Weigh Station.
See...
"Autofilling options" on page 9
User documentation for the automation
software at
"Maintenance and troubleshooting" on
page 27
"Connecting power and
15.
Pump Module User Guide
Installing a Weigh Station
www.agilent.com/chem/askb
25