Agilent Technologies Bravo Platform User Manual page 171

Hide thumbs Also See for Bravo Platform:
Table of Contents

Advertisement

A Accessories and platepads
Setting up thermal and shaking stations (Inheco controller)
Configuring an Inheco station
160
Figure
Inheco STC and MTC Controllers
To configure an Inheco station in Bravo Diagnostics:
1
In Bravo Diagnostics, click the Profiles tab, and verify that the correct profile is
initialized.
2
Click the Configuration tab.
3
In the Location is configured as list, select Accessory. The Accessories Wizard
opens.
4
In the Accessories Wizard, select the Location of the installed accessory. In the
Choose Accessory page, select Heating Shaking Station.
5
Enter the following values in the accessory parameters page.
a
In the MTCSTC Id box, select the ID number of this device. The ID number
identifies the specific device for the VWorks software.
Note: The ID number is set using a DIP switch control on the rear panel of
the Inheco STC or MTC Control. See the Inheco user documentation for a
description of how to set the ID for each slot on the STC or MTC device.
b
Select the MTCSTC Type, either STC or MTC.
c
In the Slot Id box, select the slot number on the Inheco STC or MTC rear panel
where the accessory's cable is connected.
Note: The STC has only one slot.
6
Click Finish.
7
When the configuration message appears, click Yes to initialize the accessory and
move the teachpoint to a safe height.
8
In the Profiles tab, click Update this profile.
9
Edit the location's teachpoint for the accessory. See
shaking station" on page
162.
"Teaching a thermal or thermal
Bravo Platform User Guide