E Diagnostics quick reference
Pin Tool task parameters
Related information
270
Parameter
Tip touch retract distance
(mm)
Tip touch horizontal offset
(mm)
For information about...
Opening Diagnostics
Editing the labware classes, liquid
library, or pipetting techniques
Performing a diagnostic task
Reporting a problem
Description
The vertical distance for the pins to move before
moving laterally within the well,
where
0 is the vertical distance equal to the well
bottom
> 0 is the vertical distance the pins rise above
the bottom
< 0 is the vertical distance the pins attempt to
move past the well bottom
The horizontal distance that the pins move. The
value is based on the well diameter specified by
the labware definition,
where
0 is a distance equal to the well radius
> 0 is the distance the pins attempt to move
past the well radius, which results in a more
forceful tip touch
< 0 is a distance less than the radius of the
well, resulting in a lighter tip touch or no tip
touch
See...
"Opening Bravo Diagnostics" on page 24
VWorks Automation Control Setup Guide
"Performing a task using Bravo Diagnostics"
on page 100
"Reporting problems" on page 87
Bravo Platform User Guide