5 Using diagnostics software
Jogging the liquid-handling head
Related information
94
4
Monitor the current position of the liquid-handling head by checking the W-Axis,
X-Axis, Y-Axis, and Z-Axis displays.
When the given axis is at the home position, a value of 0.00 displays.
When the liquid-handling head is not in the home position, the position is displayed
as the distance from the home position for each axis.
Note: The teachpoint coordinates for the selected location are displayed under the
Location list.
For information about...
Teachpoints
Opening diagnostics
Jog/Teach tab
Moving the liquid-handling head
manually
Setting up the Bravo Platform
See...
"Setting teachpoints" on page 34
"Opening Bravo Diagnostics" on page 24
"Jog/Teach tab quick reference" on
page 232
"Manually moving the head or gripper" on
page 77
"Workflow for setting up the Bravo
Platform" on page 16