3.1.2 Safety measures
TM safety measures fully comply with safety measures presented
in OM 2.3.1.
3.1.3 TM procedure of the device
3.1.3.1 External examination
Examination of the device should be performed in the following
order:
а) check the condition of the device's surfaces, the integrity of
seals, absence of scratches, traces of corrosion, and surface damage
of coating;
b) check the condition of the USB socket contacts.
3.1.3.2 Delivery kit completeness check
Check if the delivery kit is complete according to Table 1.2.
3.1.3.3 Operability check of the device
3.1.3.3.1 Operability check of the device and its procedure are
performed according to OM 2.3.3.
3.1.3.4 Storage battery status control
The device's battery status control is performed during daily use,
and before the long-term storage of the device. Follow the procedure
below:
- Switch on the device;
- Monitor the battery status by the indicator on the display;
- If the battery is not full, charge it by plugging the charger in the
USB connector.