Depending on the current position of the cursor in the "Cursor"
field, information about the channel, energy (in keV) and the number
of pulses will be displayed.
In addition, there is "Lin/Log" checkbox over the field of detailed
spectra viewing, which allows you to switch between spectra display
scale from linear to logarithmic, or vice versa. To the left of the
"Cursor" field you may see the ambient dose equivalent rate and the
number of pulses per second per spectrum saving time, as well as its
total accumulation time. Below the field of detailed spectra viewing,
there is information about energy calibration.
If you need to sort the loaded spectra, you may set the time
interval in the field of filtering by date to display spectra, which
creation date falls within this interval. You should check the box
opposite fields "from" and "to" and set the desired time interval.
After selecting the time interval, click "Apply" to display spectra in
accordance with the specified parameters.
"Delete spectra" button is designed to delete all saved spectra
from the device's memory.
IMPORTANT! If spectra were not saved on the PC's hard drive
(or removable disc drive), they are not subject to recovery after
removal from the device's memory.
In the upper right corner of the "Spectra" tab, there is a
button that allows exporting information about this spectrum in .xml
format in accordance with ANSI 42.42 requirements.