ECOTEST SPECTRA MKS-11GN Operating Manual page 53

Search dosimeter-radiometer
Table of Contents

Advertisement

• Enter password - enter the current password in this field;
• Enter new password - enter a new password in this field;
• Repeat new password - reenter the new password in
this field.
2.3.3.8 Energy calibration
Calibration of the device by the energy allows using a calibration
sample containing
device's detector. It is performed every 6 months maximum. It is
available only when the device is switched on in the
"Administrator" mode.
IMPORTANT! To perform calibration, it is necessary that the
device remained in regular steady-state condition for at least 2 hours
prior to calibration and the conditions must be unchanged in its
process. One also needs to make sure in the absence of other sources
of photon-ionizing radiation affecting the calibration result.
Figure 29 – Password change
232
Th radionuclide, compensate for aging of the