1.2.9 The false alarm rate for photon-ionizing radiation and
neutron rays is not more than 1 per 10 hours of operation in a stable
background environment and at the following values of threshold
levels:
- Search – 5;
- Safety – 1 µSv/h;
- Neutron – 5.
1.2.10 The device does not give false signals about the presence
of neutron radiation when exposed to photon-ionizing radiation from
60
137
Со or
Cs source at DER value of up to 100 µSv/h.
1.2.11 The device can automatically record event log files in the
nonvolatile memory, namely:
- Switch the dosimeter on;
- Switch the dosimeter off;
- Search threshold level being exceeded;
- Safety threshold level being exceeded;
- Conditional threshold level being exceeded in the detector of
neutron radiation;
- Accumulated dose threshold level being exceeded;
- Save measurement by the user's command;
- Change settings by the administrator;
- Clear the device's flash memory;
- Identification results.
1.2.12 The nonvolatile memory stores up to 20,000 records of
registered event, as well as 250 full gamma spectra.
1.2.13 The device can identify the following gamma
radionuclides:
medical radionuclides:
99m
industrial radionuclides:
152
Eu,
103
111
Tc,
Pd,
In,
22
241
Na,
Am;
18
67
F,
Ga,
123
125
131
153
I,
I,
I,
57
60
Co,
Co,
10
51
75
89
Cr,
Se,
Sr,
Хе;
201
133
Sm,
Tl,
133
137
Ba,
Cs,
99
Mo,
192
Ir,

Advertisement

Table of Contents
loading

Table of Contents