Set up and Configuration of CE in Empower 3
3
PreConfiguration of CE in Empower
For the following configuration options the PreConfiguration Utility is
required to access the CE configuration in the CE configuration editor:
Selecting PSI as pressure unit, the CE status dashboard shows the pressure in PSI while
NOTE
the online plot and the resulting Aux Traces on the Chromatogram offer the values in
mbar/bar.
6
In the New Project Wizard > Options window make sure to enable CE/CIA.
Figure 2
Enabled Options for the project
Defining the Pressure Unit (see also chapter Known Limitations on page
33)
Defining the Temperature Control mode (Temperature Control switched on
by default)
Defining the Analog In Signal (see also chapter Known Limitations on
page 33)