Example: Multiple Injections Using The Iq Kit For Ce - Agilent Technologies 7100 Technical Note

Capillary electrophoresis system in empower
Hide thumbs Also See for 7100:
Table of Contents

Advertisement

Example: Multiple injections using the IQ Kit for CE

For testing multiple injections in one sequence, the instrument method and
the experimental setup described in section Example: Single test run can be
used. See the sequence table (Figure 27) and the overlay of the resulting
electropherograms (Figure 28).
A set of six injections was used to calculate system precision. The peak
areas and relative standard deviations (Excel: STDEV) for the areas,
migration times and the corrected areas are summarized in the following
table.
Table 2
Injection 1
Injection 2
Injection 3
Injection 4
Injection 5
Injection 6
Average
Deviation STDEV:
Sixfold sample injection
Figure 27
Agilent 7100 CE in Empower Environment
Peak areas and relative standard deviations
Area
50049
50547
50189
49767
49187
49579
49886
0.96 %
Sequence table for multiple injections
t
[min]
t
[s]
m
m
2.730
163.80
2.723
163.38
2.717
163.02
2.711
162.66
2.706
162.36
2.702
162.12
2.715
162.89
0.39 %
0.39 %
APPENDIX - Examples
corr. Area (area/t
[s])
m
3.05549E+02
3.09383E+02
3.07870E+02
3.05957E+02
3.02950E+02
3.05817E+02
3.06254E+02
0.72 %
39