Sequence Execution/Sample sets
Empower offers multiple injections from the same vial as well as
consecutive injections from different injection vials. The details have to be
added to the sample/sample set table (sequence), depending on the
required workflow.
1
2
The parameter Inj. Vol (µl) requires a value. This value is not used for the injection, but it
NOTE
is reported. The instrument method parameters (for example: pressure and time) are used
for the injection.
(Instead of volume the quantifiable parameters are pressure × time for hydrodynamic
injection, or voltage × time for electrokinetic injection).
Entering an injection volume of 0 µl does not result in a blank run. The value is ignored for
the run execution, but it is reported.
3
Agilent 7100 CE in Empower Environment
Click the Sample tab to open the table.
Click into the first line to fill in the appropriate values. For the
following values the entries are mandatory:
Vial position
Injection volume
Number of injections
Sample name
Method set
Run time (use the same time as in the instrument method)
Start the sequence. Reminder: During run execution, the red Stop
button allows you to stop the run at any time.
Using the CE in Empower
23