Using the CE in Empower
16
d
General tab: Offers option to mark this method as shutdown
method, see page 20. Allows selective shutdown of the CE and
DAD.
e
Instrument Configuration tab: The CE system configuration is
available in the Instrument Configuration tab. All settings in this
dialog must be defined using the PreConfiguration Utility. Changes
in this section of the method are not applied to the system; they are
saved with the instrument method only. Any changes, such as
enabling temperature control, need to be done in the
PreConfiguration Utility to apply the new configuration to the
physical instrument. In addition, the LAC/E box has to be rebooted
and a new autoconfiguration is required in order to adjust the new
configuration with the LC Driver and Empower.
Agilent 7100 CE in Empower Environment