Set up and Configuration of CE in Empower 3
Configuration of CE in Empower
Refer to the Waters Empower documentation for installation and
configuration of the LC/CE system in Empower.
1
2
3
4
10
Open the Waters DHCP Server Configuration window by entering the
following command in the windows command line:
C:\Empower\Instruments\Waters DHCP Server Configuration.exe.
OR
In the Empower Configuration Manager window, select Node in the tree
on the left-hand side, then right-click on the node you want to add the
instrument to and select Properties. Select the Configure DHCP tab and
click Configure DHCP.
Add the IP Address and MAC Address manually for the CE mainframe.
Select Instrument Type AgilentLC and click OK to leave the screen.
Figure 7
Entering communication details for CE
Access the Nodes Properties in the Empower Configuration Manager and
verify that the Instrument is shown as OK.
Agilent 7100 CE in Empower Environment