2. Selecting the RS-232C Baud Rate

The RS-232C baud rate can be changed according to the setting made on the computer that is used for remote
measurement.
[Operating Procedure]
1
.
Press the
key.
The LCD display section will switch to the menu selection
screen.
2
.
Press the
key to open the RS232C baud rate
selection screen.
Each time the
key is pressed, the screen will switch in the
order PROBE → SYNC → ID Name input → RANGE →
Measurement Speed → Number of Digits → RS232C Baud
Rate → PROBE.
3
.
Press the
key until the desired baud rate ap-
pears.
Each time the
key is pressed, the baud rate switches in
the order 9600 → 4800 → 2400 → 1200 → 600 → 300 →
38400 → 19200→ 9600.
4
.
Press the
key to confirm the selection.
* By default (factory setting), the instrument is set so that [9600] will be selected automatically when the POWER switch is set to ON.
* To cancel selection of RS-232C baud rate, press the
<Notes when Selecting the RS-232C Baud Rate>
● The specified RS-232C baud rate will be kept even if the POWER switch is set to OFF. The selected RS-232C
baud rate will be effective when the POWER switch is set to ON.
<Reference>
Communication parameter setting (RS-232C)
Set the same communications settings as those on the instrument to the computer.
Baud rate: 38400, 19200, 9600 (factory setting), 4800, 2400, 1200, 600, 300, BPS
Start bit: 1 bit
Character length: 7 bits (ASCII code)
Parity check: EVEN
Stop bit: 2 bits
PQRS
TUV
WXYZ
1
7
8
9
MENU
GHI
JKL
MNO
LOCK
4
5
6
ALPHA
White
ABC
DEF
- SPACE
2
1
2
3
CAL
3
Red
Green
Blue
0
.
4
ENTER
key.
85
Menu selection screen
MENU : SELECT
PUSH SPACE KEY
RS-232C baud rate selection screen
SELECT : BAUD
38400
SELECT : BAUD
19200
SELECT : BAUD
9600
Press the
key until the
desired baud
rate appears.