[Setting Procedure]
1
.
Press the MODE key to select the measurement mode
for which you want to set the range.
2
.
Press the
The LCD display section will switch to the menu selection screen.
3
.
Press the
Each time the
PROBE → SYNC → ID Name input → RANGE → Measurement
Speed → Number of Digits → RS232C Baud Rate → PROBE.
4
.
Enter the desired range value.
Use the ten-key (
to the right each time a value is entered.)
Each time the
Lv, between G and B/G, R/G or between R and B/G, R/G.
In this example, the "x, y" range is set to 2.5%, and the "Lv" range is
set to 2.0%.
1 Press the
,
2 Press the
The cursor (_) will move to the "Lv" position.
3 Press the
,
5
.
Press the
The ranges will be set.
* To cancel range setting, press the
* By default (factory setting), the ranges are set to 10%.
<Error Messages in LCD Display Section>
● "E4" (after the
key is pressed)
• Cause : 0.0% was entered.
• Action : Enter a correct value and then press the
key.
key to open the RANGE setting screen.
key is pressed, the screen will switch in the order
to
,
) to enter the value. (The cursor moves
key is pressed, the cursor moves between x, y and
and then
key to set the "x, y" range.
key.
and then
key to set the "Lv" range.
key.
key before pressing the
1
4
key.
... For other error messages, refer to page 107.
key. The settable range is from 0.1 to 99%.
70
2
3
5
Menu selection screen
MENU : SELECT
PUSH SPACE KEY
Range setting screen
(For xyLv, T∆uvLv, u'v'Lv
or XYZ mode)
RANGE x,y
Lv
(%) 10
10
_
For analyzer mode (G reference)
RANGE G
B/G,R/G
(%) 10
10
_
For analyzer mode (R reference)
RANGE R
B/R,G/R
(%) 10
10
ANGE x,y Lv
(%) 2.5 10
ANGE x,y Lv
(%) 2.5 2.0
ANGE x,y Lv
E4(%) 0.0 10