Internal Clock and Calendar
The internal clock is set the first time you use the Setup command from the MultiLab
software to program the MultiLogPRO. The clock is automatically updated to the
computer's time and date whenever you connect your MultiLogPRO to a computer
and perform Setup in MultiLab.
When you shut off the MultiLogPRO, the internal clock and calendar is not saved. To
restore the clock and calendar, you must reconnect the MultiLogPRO to the
computer and Setup the logger again.
Clear the Memory
If you want to start recording and the MultiLogPRO's internal memory is full you will
see this message at the bottom of the display:
Press the Forward arrow button
In order to clear the MultiLogPRO's memory when it is not full, use the Memory
clear command from the Configuration menu (see page 21), or clear the memory
from the Logger menu in the MultiLab software (see page Error! Bookmark not
defined.).
Choose the Right Setup
1. Sampling rate - The sampling rate should be determined by the frequency of
the phenomenon being sampled. If the phenomenon is periodic, sample at a
rate of at least twice the expected frequency. For example, sound recordings
should be sampled at the highest sampling rate – 20,800/sec, but changes in
room temperature can be measured at slower rates such as once per second or
even slower, depending on the speed of the expected changes.
Fourier Systems Ltd.
16 Hamelacha St., POB 11681, Rosh Ha'ayin 48091,
Tel: +972-3-901-4849, Fax: +972-3-901-4999
www.FourierEdu.com
In–1: Microphone
In–2: Light
In–3: Empty
In–4: Empty
SAMPLES = 200
DISPLAY = graphic
M e m f u l l , c l e a r = ( > )
to clear the memory