Batch Print Mode - Hydronix Hydro-View HV02 User Manual

Table of Contents

Advertisement

BATCH PRINT MODE

AAA.AA
<CRLF>
Batch print mode
The Hydro-View provides a report at the end of a batch (as determined by the state of the
average/hold input) on the relevant port which may be printed or captured by an external
device.
This data stream is in ASCII format and contains the NOW and AVERAGE moisture
values updated once per second.
The data format used is as follows...
Now = NNN.NN%, Avg = AAA.AA%<CRLF>
where...
NNN.NN
AAA.AA
<CRLF>
The data rate/format used is 9600 baud, 8 data bits, no parity, 1 stop bit. This cannot be
altered.
The batch print format changes when the unit is configured as a Master on the other serial
port. The Hydro-View provides a report at the end of a batch as determined by the state of
the average/hold inputs across all units involved in the group.
The data format used is as follows...
M1 = MMM.MM%, M2 = NNN.NN%, Avg = AAA.AA%<CRLF>
where...
MMM.MM
NNN.NN
AAA.AA
<CRLF>
74
HYDRO-VIEW USER GUIDE D0022 REV 1.06
is the group average moisture value in the range -99.99 to 99.99%.
is the ASCII carriage return - line feed combination (hex 0D, hex 0A).
is the NOW moisture value in the range -99.99
to 99.99%.
is the AVERAGE moisture value in the range -
99.99 to 99.99%.
is the ASCII carriage return - line feed
combination (hex 0D, hex 0A).
is the group M1 value in the range -99.99 to 99.99%.
is the group M2 value in the range -99.99 to 99.99%.
is the AVERAGE moisture value in the range -99.99 to 99.99%.
is the ASCII carriage return - line feed combination (hex 0D, hex 0A).