Peristaltic Pump Installation - Thermo Scientific Dionex Chemically Regenerated Suppressor 500 Series Product Manual

Hide thumbs Also See for Dionex Chemically Regenerated Suppressor 500 Series:
Table of Contents

Advertisement

2
– Installation
B. Make the following liquid line connections. See Figure 8, "Configuration of the
C. Fill the regenerant reservoir with the recommended regenerant (Refer to Tables 3
D. Adjust the regenerant flow rate to approximately 5–10 mL/min. for the Dionex
!
A safety relief valve on the reservoir regulator prevents pressure greater than 25 psi from
being applied to the reservoir.
NOTE

2.5 Peristaltic Pump installation

For detailed installation instructions please refer to the MASTERFLEX
Quick Start Guide (Document No. 065203)
Thermo Scientific
031727-07
External Chemical Mode with the Dionex CRS 500 Suppressor."
1.
Use a coupler (P/N 039056) to connect one end of the 30" tubing assembly (P/N
035727) that comes in the Installation Kit to the water reservoir. Connect the other
end of this tubing to the REGEN IN port of the Dionex Chemically Regenerated
Suppressor.
2.
Using a coupler (P/N 039056) and a 1/8" o.d. piece of tubing (P/N 035728) from
the Installation Kit, connect one end of this tubing to the REGEN OUT port and
direct the other end to waste.
or 4, "Matching the Regenerant concentration and flow rate to eluent
concentration and flow rate"). Make sure that the O-ring is inside the cap of the
reservoir before screwing the cap onto the reservoir. Screw the cap onto the reservoir
tightly and place the reservoir near the Chromatography Module.
CRS 500 4 mm and 5–8 mL/min for the Dionex CRS 500 2 mm.
Product Manual for Dionex CRS 500 Suppressor
®
®
C/L
Peristaltic Pump
Page 24 of 40