Technical Specification; General Data; Functions - ILLKO REVEXplus USB User Manual

Table of Contents

Advertisement

User's Manual

6. TECHNICAL SPECIFICATION

6.1. General data

Protective class .......... II (double insulation)
Over voltage class ............................. CAT II
Pollution degree .......................................... 2
Degree of protection ............................. IP 20
Reference condition:
Line voltage . 230V1%/50Hz,60Hz 1%
Ambient temperature ............. (23  2) C
Humidity ............ 45  55 % RH @ 23 C
Position ...................................... arbitrary
Operating condition:
Line voltage 230V10%/50Hz,60Hz 1%
Ambient temperature ............ 0C  40C

6.2. Functions

Earth bond 200 mA
Entire measuring range
Measuring sub range
Resolution
Reference error
Operating error
Measuring sub range
Resolution
Reference error
Operating error
Operating range of use
(according to EN 61557-4)
Open-circuit voltage
Test current (R ≤ 2 )
Test lead compensation
PE current during operation
Entire measuring range
Measuring sub range
Resolution
Reference error
Operating error
Measuring sub range
Resolution
Reference error
Operating error
Test voltage
Frequency characteristic
of internal test circuitry
Functionality test
Copyright © 2006–2019 ILLKO, s.r.o.
0.000 ÷ 20.00 
0.000 ÷ 0.999 
0.001 
(1 % of R + 5 D)
(2 % of R + 5 D)
1.00 ÷ 20.00 
0.01 
(1 % of R + 3 D)
(2 % of R + 3 D)
0.018 ÷ 20.00 
AC 5 ÷ 12 V
≥ AC 200 mA
yes (up to 2 )
0.000 ÷ 10.00 mA
0.000 ÷ 1.499 mA
0.001 mA
(1 % of R + 5 D)
(2 % of R + 8 D)
1.50 ÷ 10.00 mA
0.01 mA
(1 % of R + 5 D)
(2 % of R + 5 D)
line voltage
according to EN 61010
0.090 ÷ 0.110 mA
Humidity ............ max. 75 % RH @ 23C
(condensation not allowed)
Elevation..............................max. 2000 m
Max. supply current .............................. 16 A
(it depends on tested appliance)
Quiescent current ............................  50 mA
Dimensions (w x d x h) . 150 x 195 x 50 mm
Mass (without accessories) .................... 1 kg
Storage temperature / storage humidity.........
.......................-10  50C / max. 75% RH
Case resistance to temperature ............. 70C
PE test ........................................... automatic
PC connection (REVEXplus USB only).USB
Insulation resistance
Entire measuring range
Measuring sub range
Resolution
Reference error
Operating error
Measuring sub range
Resolution
Reference error
Operating error
Measuring sub range
Resolution
Reference error
Operating error
Measuring sub range
Resolution
Reference error
Operating error
Operating range of use
(accord. to EN 61557-2)
Nominal output voltage Un DC 500 V
Open-circuit voltage
Nominal test current
Short-circuit current
Functionality test
Mains voltage U
Measuring range
Resolution
Reference error
Operating error
34
REVEXplus, REVEXplus USB
0.000 k ÷ 100.0 M
0.0 k ÷ 999.9 k
0.1 k
(2 % of R + 8 D)
(3 % of R + 8 D)
1.000 M ÷ 3.999 M
0.001 M
(2 % of R + 8 D)
(3 % of R + 8 D)
4.00 M ÷ 19.99 M
0.01 M
(2 % of R + 8 D)
(3 % of R + 8 D)
20.0 M ÷ 100.0 M
0.1 M
(4 % of R + 15 D)
(5 % of R + 15 D)
3.0 k ÷ 100.0 M
(-0%, + 50%) Un
≥ DC 1 mA
< DC 10 mA
2.000 ÷ 2.600 M
L-N
207.0 ÷ 253.0 V
0.1 V
(1 % of R + 2 D)
(2 % of R + 2 D)

Advertisement

Table of Contents
loading

Table of Contents