• Locking of measurement – press the START key and then press the LOCK key.
Release both keys. Measurement is in progress and current measured values are
displayed until the START key is again pressed and released.
Measuring process is indicated by (START) indicator.

4.1.2. HOLD function

When measurement is in progress then current measured values are displayed. By press and
release of the HOLD key the last measured result is displayed on display even if
measurement either ended or continues. The HOLD function is indicated by such way that
blinking of the (START) indicator changes to unremitting light.
HOLD function can be cancelled (measurement is still in progress) by press and release of the
HOLD key. The (START) indicator then blinks again.
4.2. Earth bond 200 mA - R
Make sure that tested equipment is deenergized – its mains voltage must be
disconnected before starting the measurement!
Do not connect any external voltage between R
test socket 1 or PE terminal 4.
If you are going to test equipment which can be charged to voltage > 25V (e.g.
filter capacitors), discharge it before test!
In case that external voltage is present on appliance under test, the
(PROTECT) indicator turns on and „Prot" message is displayed. Such status
is dangerous fault condition; disconnect equipment under test from the
REVEXplus immediately to avoid its damage!

4.2.1. Test lead resistance compensation

Compensation means that the test lead resistance is measured and then automatically
memorized and subtracted from all R
is thus displayed. The compensation constant remains in instrument's memory even if
the instrument is switched off. And so the lead compensation is valid until new
compensation is carried out.
1
2
3
Copyright © 2006–2019 ILLKO, s.r.o.
PE
M
R
A
m
PE
W
Cal
LOC K
ST ART
terminal 3 and either PE in
PE
measurements. The real earth bond resistance
PE
PE
K
A
V
VA
cos
Fig. 3 – Test lead resistance compensation
14
REVEXplus, REVEXplus USB
R
I
PE
D

Advertisement

Table of Contents
loading

Table of Contents