Mains Voltage Measurement; Current Consumption Measurement - ILLKO REVEXplus USB User Manual

Table of Contents

Advertisement

User's Manual

4.7.1. Mains voltage measurement

Press the U, I key and release it repetitively in order to select voltage measurement
function U. The (V) indicator has to light up.
Press the START key and release it. Measurement is indicated by the (START)
indicator.
After finishing of measurement (the (START) indicator turns off) the last measured
value of mains voltage is displayed.
1
2

4.7.2. Current consumption measurement

WARNING - DANGER!
If test voltage is present in Test socket 1 – it is indicated by the (START)
indicator, then tested equipment starts to run after its mains switch is
switched on! Take into consideration all safety warnings, referred to
equipment under test!
Do not touch exposed conductive parts of tested object after mains switch is
switched on!
Test Socket 1 of the REVEXplus must not be used for continuous power
supply of appliance! It is intended for test purposes only; max. test duration
is 60 s @ 16A.
1
2
Connect tested appliance
3
Switch on mains
4
switch of tested
appliance
Copyright © 2006–2019 ILLKO, s.r.o.
U,I
ST ART
U,I
ST ART
M
K
A
A
V
m
W
VA
cos
Fig. 13 – Mains voltage measurement
M
K
A
A
V
m
W
VA
cos
Fig. 14 – Current consumption measurement
26
REVEXplus, REVEXplus USB
Function selection
Measurement
Function selection
Connect tested appliance
to test socket 1
Test socket 1 energizing
Measurement