User's Manual
4.7.1. Mains voltage measurement
Press the U, I key and release it repetitively in order to select voltage measurement
•
function U. The (V) indicator has to light up.
Press the START key and release it. Measurement is indicated by the (START)
•
indicator.
•
After finishing of measurement (the (START) indicator turns off) the last measured
value of mains voltage is displayed.
1
2
4.7.2. Current consumption measurement
WARNING - DANGER!
•
If test voltage is present in Test socket 1 – it is indicated by the (START)
indicator, then tested equipment starts to run after its mains switch is
switched on! Take into consideration all safety warnings, referred to
equipment under test!
•
Do not touch exposed conductive parts of tested object after mains switch is
switched on!
•
Test Socket 1 of the REVEXplus must not be used for continuous power
supply of appliance! It is intended for test purposes only; max. test duration
is 60 s @ 16A.
1
2
Connect tested appliance
3
Switch on mains
4
switch of tested
appliance
Copyright © 2006–2019 ILLKO, s.r.o.
U,I
ST ART
U,I
ST ART
M
K
A
A
V
m
W
VA
cos
Fig. 13 – Mains voltage measurement
M
K
A
A
V
m
W
VA
cos
Fig. 14 – Current consumption measurement
26
REVEXplus, REVEXplus USB
Function selection
Measurement
Function selection
Connect tested appliance
to test socket 1
Test socket 1 energizing
Measurement