User's Manual
2.2. Terminals
•
Do not connect any external voltage to any terminal or connector. The only
exception is power supply cord 5 which is intended for powering of the
REVEXplus from 230 V / 50 ÷ 60 Hz mains outlet.
•
Max. load on test socket 1 is 16 A, max test duration is 60 s.
•
Use original accessories only.
Test Socket for connection of tested appliance's power supply cord.
1
– touch leakage current test terminal.
2
I
D
– earth bond test terminal.
3
R
PE
PE – terminal for connecting of exposed conductive part of tested appliance (the
4
terminal is connected to PE terminal of Test Socket 1).
5
Power supply cord.
6
R
+I
ISO
SUB
EXT INPUT – current clamp / three-phase adapter / other adapters connector.
7
8
USB connector (REVEXplus USB only).
2.3. Keyboard and indicators
Keyboard – each key selects one or more functions
R
– Earth bond test.
9
PE
10 I
– Substitute leakage current and touch leakage current.
SUB
11 I
– PE current during operation and touch leakage current.
PE
12 I
– Differential leakage current and touch leakage current.
13 I
– Touch leakage current on protective class I appliances.
D
14 R
– Insulation resistance.
ISO
15 U,I – Mains voltage and appliance's current.
16 EXT – EXT INPUT for current clamp / three-phase adapter activation.
17 P,S – Active power / apparent power / cos φ.
HOLD – Holding of displayed data.
18 CAL – Test lead resistance compensation on R
LOCK - Locks the START key.
19 START – Starts measurement.
Indicators
20 LED display.
(M) − Insulation resistance in
(k) − Insulation resistance in k
() − Earth bond resistance in
24 (mA) − Leakage or touch current in mA.
25 (A) − Supply current of tested application in A.
26 (V) − Mains voltage in V.
27 (W) − Active power in W.
28 (VA) − Apparent power in VA.
29 (cos φ) − cos φ.
Copyright © 2006–2019 ILLKO, s.r.o.
– insulation resistance / substitute leakage current test terminal.
REVEXplus, REVEXplus USB
function / basic functionality test.
PE
8