Utilization Of Other External Adapters - Ext - ILLKO REVEXplus USB User Manual

Table of Contents

Advertisement

User's Manual
Connect optional current clamp to the EXT INPUT terminal on the rear side of the
REVEXplus.
Press and release the P, S key repetitively in order to select required function. The
(W), (VA) or (cos ) indicator has to light up and (EXT) indicator has to light up, too.
Switch on tested appliance's mains switch and put it into operation.
Enclose with current clamp's jaws live conductor L of measured appliance in a way that
arrow on current clamp points from power source to appliance.
Press the START key; measurement starts to run (see chapter 4.1. for measurement
control details). Measurement is indicated by the (START) indicator.
Finish measurement after displayed result is stabilized (see chapter 4.1. for
measurement control details).
After finishing of measurement (the (START) indicator turns off) the last measured
value is displayed. You can check other measured quantities – press and release the P,
S key repetitively.
Switch off tested appliance's mains switch.
4.9. Utilization of other external adapters – EXT
By repetitive press and release of the EXT key it is possible to select besides other functions
the function, where both the (V) and (EXT) indicators are turned on.
This function is intended for connecting of other adapters, e.g. the WELDtest - adapter
enabling measuring of arc welding equipments output voltage in accordance with
EN 60974-4.
• The way of connecting of adapter to instrument, description of measuring
procedure and other necessary information are published in adapter's user
manual.
• The
EXT
(EXT) indicator lights up and test socket 1 is not functional.
It is prohibited to connect any external voltage to the EXT INPUT terminal!
Use original accessories only!