User's Manual
Press the U, I key and release it repetitively in order to select current consumption
•
measurement function I. The (A) indicator has to light up.
•
Connect tested appliance to test socket 1.
Press the START key; measurement starts to run (see chapter 4.1. for measurement
•
control details). Measurement is indicated by the (START) indicator.
•
Switch on tested appliance's mains switch and put it into operation.
•
Finish measurement after displayed result is stabilized (see chapter 4.1. for
measurement control details).
•
After finishing of measurement (the (START) indicator turns off) the last measured
value of tested appliance's supply current is displayed.
•
Switch off tested appliance's mains switch.
4.7.3. Current measurement by means of current clamp -
The
EXT
key activates external input EXT INPUT. If it is activated, the (EXT)
indicator lights up and test socket 1 is not functional!
•
Connect optional current clamp to the EXT INPUT terminal on the rear side of the
REVEXplus.
Press and release the EXT key repetitively until the (A) indicator lights up. The
•
(EXT) indicator is turned on, too.
•
Switch on tested appliance's mains switch and put it into operation.
•
Enclose with current clamp's jaws live conductor L of measured appliance.
Press the START key; measurement starts to run (see chapter 4.1. for measurement
•
control details). Measurement is indicated by the (START) indicator.
•
Finish measurement after displayed result is stabilized (see chapter 4.1. for
measurement control details).
•
After finishing of measurement (the (START) indicator turns off) the last measured
value of current is displayed.
•
Switch off tested appliance's mains switch.
1
Connect current clamp to the instrument and
2
enclose with its jaws live conductor of
measured appliance.
3
Switch on mains switch of tested appliance
4
Copyright © 2006–2019 ILLKO, s.r.o.
I NV
I
EXT
EXT
PROTEC T
ST ART
Fig. 15 – Current measurement by means of current clamp
EXT
M
K
A
A
V
m
W
V A
cos
27
REVEXplus, REVEXplus USB
Function selection
Connect current clamp
Measurement