User's Manual
Step 1 - preparation for measurement
•
Select the I
SUB
and indicators (I
•
Connect tested appliance:
One-phase class II appliance with movable power supply cord – plug in its
-
mains plug to test socket 1.
•
Connect test lead P2011 + test tip P3011 to the instrument's ID terminal.
•
Switch on tested appliance's mains switch.
Step 2 – measurement
•
Press test tip against exposed conductive part of tested appliance.
Press the START key; measurement starts to run (see chapter 4.1. for measurement
•
control details). Measurement is indicated by the (START) indicator.
•
After finishing of measurement (the (START) indicator turns off) the last measured
value of substitute touch leakage current is displayed.
4.5. PE current during operation - I
I
4.5.1. Protective class I one-phase appliances
WARNING - DANGER!
•
If test voltage is present in Test socket 1 – it is indicated by the (START)
indicator, then tested equipment starts to run after its mains switch is
switched on! Take into consideration all safety warnings, referred to
equipment under test!
•
Do not touch exposed conductive parts of tested object after mains switch is
switched on!
•
Firstly earth bond resistance (chapter 4.2.) must be tested; it is recommended
to test insulation resistance if it is possible, too (chapter 4.3.). Then you can
execute leakage current measurement!
•
Equipment under test must be isolated from ground (from the earth
potential) during measurement of I
•
Test Socket 1 of the REVEXplus must not be used for continuous power
supply of appliance! It is intended for test purposes only; max. test duration
is 60 s @ 16A.
The REVEXplus is equipped with electronic checking of excessive leakage current
during measurement. If leakage current exceeds limit value about 12 mA, then mains
voltage is disconnected from test socket 1. This status is indicated by displaying
„Prot" message; the indicator (PROTECT) turns on, too.
1
2
Copyright © 2006–2019 ILLKO, s.r.o.
function: pres the I
SUB
) and (mA) light up.
SUB
INV
EXT
I
PE
PRO TEC T
Connect tested appliance
key and release it. Symbol „- -" is displayed
and differential leakage current -
PE
!
PE
M
K
A
A
V
m
W
VA
cos
19
REVEXplus, REVEXplus USB
Function selection
Connect tested appliance to
test socket 1