Setting The Dip Switches (Only For Zone 2/22 And Div 2); Ranges / Maximum Cable Length Of Connected Cable From The Universal Supply Module To The Host Or Pc - Bartec BCS 3600ex Series User Manual

Hide thumbs Also See for BCS 3600ex Series:
Table of Contents

Advertisement

4 Product description
4.6.2
4.6.3
36 / 106

Setting the DIP switches (only for Zone 2/22 and Div 2)

The DIP switches (X18) must be set as follows depending on the interface used
Figure 14 DIP switches and switch positions
Setting Interface (Dipp-switch S1, S2 and S3)
Interface
S1
RS 232
0
RS 422
0
RS 485
0
USB-SPP
1
Settings USM Version (Dipp-switch S4)
Version
S1
corded
-
Bluetooth
-
Table 33 Setting
Ranges / maximum cable length of connected cable from the universal supply
module to the host or PC
Interface
RS 232
RS 422
RS 485
USB-SPP
Table 34 Ranges / maximum cable length of connected cable
S2
S3
0
0
1
1
0
1
1
1
S2
S3
-
-
-
-
Range
15 m
1000 m
1200 m
5 m
Subject to technical modifications
Handscanner
BCS 3600
series
ex
S4
-
-
-
-
S4
0
1
50 ft.
3280 ft.
3937 ft.
16 ft.
Revision A / 04/2019