How To Update Firmware/Software Using Web Ui - Viavi mA-3011 Operation Manual

Axie embedded host module
Hide thumbs Also See for mA-3011:
Table of Contents

Advertisement

Chapter 5 Care and Maintenance
Maintaining the device

How to Update Firmware/Software using Web UI

This procedure is used to update or install firmware/software in an mA-301x Module
when the module is installed in an AXIe Chassis that is not part of the CMP AXIe
Chassis product family.
1
Download the software .tar.gz or .tgz file from VIAVI Solutions website to a local
directory.
2
Extract contents of .tar.gz or .tar file local directory.
3
Burn tar file to CD.
4
Connect external CD-ROM device to mA-3011 Embedded Host module, via a
Y-Cable or cable with 2 USB connectors, and then insert CD-ROM into external
CD-ROM drive.
NOTE
A Y-cable or dual USB connectors is required to draw enough current to power external
CD-ROM device.
5
Navigate to System Update Page.
6
Change Update Mode to File Path.
7
Update file path to point to CD-ROM.
8
Press Check for Update Button. Wait while mA-301x Shelf Manager queries the
software location.
9
If a more current version of software is identified, the Update Button is enabled.
Press the Update Button.
10 Wait while the mA-301x Shelf Manager performs the update. Do not interrupt this
process.
NOTE
When the update is complete disconnect the CD-ROM device.