Addendum A. oxy.IQ Safety Manual

oxy.IQ
MODEL
DWG NO.
Certification & Safety Statements for the oxy.IQ oxygen Transmitter
The oxy.IQ is a highly reliable and cost-effective two-wire, loop-powered transmitter
with a linearized 4 to 20 mA output. It measures oxygen in ten ppm ranges and
seven percentage ranges. All ranges are user-selectable. This compact transmitter
uses proven sensor technology to accurately measure O
hazardous environments.
When installing this apparatus, the following requirements must be met:
• The system is covered by the certificate numbers FM14ATEX0032X and IECEx
FMG 14.00016X as shown on the labels on the following page. The system
temperature code is T4 in the temperature range of -20 to 60°C.
• The apparatus should be de-energized before servicing.
• Installation shall be in accordance with the installation instructions and the
National Electrical Code® ANSI/NFPA 70, the Canadian Electrical Code C22.1, or
IEC/EN 60079-14, as applicable.
• Equipment is of type Intrinsically Safe and complies with: EN 60079-0:2012,
EN 60079-11:2012, IEC 60079-0:2007, IEC 60079-11:2006, IEC 60529:1992,
C22.2 No.1010.1:2004, CAN/ CSA-E60079-0:2011, CAN/CSA-E60079-11:2011,
C22.2.60529:2005, FM Class 3600:2011, FM Class 3610:2010, FM Class 3810:2005,
ANSI/ ISA 61010-1:2004 Ed.2, ANSI/ ISA 60079-0:2009 Ed.5,
ANSI/ ISA 60079-11:2009 Ed.5.
• The product contains no exposed parts which produce surface temperature
infrared, electromagnetic ionizing, or non-electrical dangers.
• The product must not be subjected to mechanical or thermal stresses in excess
of those permitted in the certification documentation and the instruction
manual.
• The product cannot be repaired by the user; it must be replaced by an equivalent
certified product. Repairs should only be carried out by the manufacturer or by
an approved repairer. Please contact GE Customer Support Center for repair,
maintenance or replacement of the product. For oxygen sensor replacements
please contact GE Customer Support Center.
U.S.A
The Boston Center
1100 Technology Park Drive
Billerica, MA 01821
U.S.A
Tel: 800 833 9438 (toll-free), 978 437 1000
E-mail:sensing@ge.com
oxy.IQ User's Manual
GE Sensing
1100 Technology Park Drive
Billerica, MA 01821
714-1344
Page 2 of 6
Addendum A. oxy.IQ Safety Manual
Oxy.IQ Safety Manual
REV. C
in a variety of gases, even in
2
Ireland
GE Sensing EMEA
Free Zone East, Shannon,
Co. Clare, V14 V992, Ireland
Tel: +35 361 470200
TITLE
72

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents