Installing An Oxygen Sensor - GE Baker Hughes oxy.IQ User Manual

Panametrics oxygen analyzer
Table of Contents

Advertisement

Chapter 2. Installation

2.3 Installing an Oxygen Sensor

To install a new or replacement oxygen sensor in the oxy.IQ, refer to
Figure 6 below and complete the following steps:
Sensor Base
Knurled Nut
Oxygen Sensor Ring
Sensor Manifold
1. Disconnect the power from the oxy.IQ.
2. Loosen the blue knurled nut and remove the oxy.IQ electronics
module from the sensor manifold. If a previous oxygen sensor is
already in place, remove and discard it.
10
Figure 6: Oxygen Sensor Installation
oxy.IQ User's Manual