GE Baker Hughes oxy.IQ User Manual page 83

Panametrics oxygen analyzer
Table of Contents

Advertisement

Addendum A. oxy.IQ Safety Manual
oxy.IQ
MODEL
DWG NO.
• Only trained, competent personnel may install, operate and maintain the
equipment.
• For product training please contact GE Customer Support Center:
U.S.A
The Boston Center
1100 Technology Park Drive
Billerica, MA 01821
U.S.A
Tel: 800 833 9438 (toll-free), 978 437 1000
E-mail:sensing@ge.com
• The product is an electrical apparatus and must be installed in the hazardous
area in accordance with the requirements of the EC Type Examination
Certificate. The installation must be carried out in accordance with all the
appropriate international, national and local standard codes and practices and
site regulations for Intrinsically Safe apparatus and in accordance with the
instructions contained in the manual. Access to the circuitry must not be made
during operation.
• The maximum operating pressure for the oxy.IQ is 10psi. Be sure the sample
conditioning system is designed to maintain the oxy.IQ pressure below these
limits, and that the oxy.IQ outlet is vented to atmosphere during operation and
calibration.
• WARNING - Substitution of components may impair Intrinsic Safety.
• WARNING - To prevent ignition of flammable or combustible atmospheres,
disconnect power before servicing.
• AVERTISSEMENT - Pour éviter l'inflammation d'atmosphères inflammables ou
combustibles, débrancher l'alimentation avant l'entretien.
• AVERTISSEMENT - Remplacement des composants peut compromettre la
sécurité intrinsèque.
• Equipment is not intended for the measurement of oxygen in fluid of liquid
phase.
Special Conditions of Safe Use
1. The Model oxy.IQ will not pass the 500V dielectric test. This must be taken into
account upon installation.
2. When installed in a Division 2 location, the Model oxy.IQ oxygen transmitter
shall be installed in a final enclosure accepting a Division 2 wiring method per
NEC/CEC.
73
GE Sensing
1100 Technology Park Drive
Billerica, MA 01821
714-1344
Page 3 of 6
Oxy.IQ Safety Manual
REV. C
Ireland
GE Sensing EMEA
Free Zone East, Shannon,
Co. Clare, V14 V992, Ireland
Tel: +35 361 470200
oxy.IQ User's Manual
TITLE

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents