GE Baker Hughes oxy.IQ User Manual page 43

Panametrics oxygen analyzer
Table of Contents

Advertisement

6.3 All Installations (cont.)
Repeatability
±1% of range
±2% of range for the 0 to 10 ppm
Resolution
±0.1% of range
Linearity
±2% of range (OX-1, 2, 3, 5)
±5% of range (OX-4)
O
Sensor Operating Temperature
2
32°F to 113°F (0°C to 45°C)
Sample Pressure
Vented to atmosphere during operation and calibration
Atmospheric Pressure Effect
±0.13% of reading per mmHg (directly proportional to absolute
pressure)
Note:
During calibration, pressure and flow must be kept constant.
Process Connections
1/8" NPT-F inlet and outlet
Dimensions
4.10 x 2.75 x 2.05 in. (104.1 x 69.9 x 52.1 mm)
Weight
1.35 lb (612 grams)
oxy.IQ User's Manual
Chapter 6. Specifications
O
range (OX-1, 2)
v
2
33

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents